Capillary Ferrules
Capillary Ferrules
Selecting proper material and type of GC connection ferrule upon the application is important. GL Sciences offers extensive line up of ferrules for various analytical scene.
![グラファイトの画像](jpg/img01-8.jpg)
Graphite Ferrule
![15 % グラファイト/85 %ベスペルの画像](jpg/img04-4.jpg)
15 % Graphite / 85 % Vespel Ferrule
![SilTite金属の画像](jpg/img03-5.jpg)
SilTite Metal Ferrule
![SilTite FingerTiteの画像](jpg/img02-16.jpg)
SilTite FingerTite Ferrule
|
Graphite
|
Super Graphite
|
15 % Graphite/ 85 % Vespel
|
SilTite Metal
|
SilTite FingerTite
|
---|---|---|---|---|---|
Mechanical Strength
|
Soft
|
Soft
|
Hard
|
Hard
|
Hard
|
Max. Operating Temp.
|
450 ºC
|
450 ºC
|
350 ºC
|
450 ºC and more
|
450 ºC and more
|
Reuse
|
Possible
|
Possible
|
Not Possible
|
Not Possible
|
Not Possible
|
Strength of Tightening
|
Week
|
Week
|
Middle
|
Strong
|
Strong
|
Retighntening
|
No Need
|
No Need
|
Need
|
No Need
|
No Need
|
GC-MS
|
Not Available
|
Not Available
|
Available
|
Available
|
Available
|
Original Conditioning
|
-
|
Yes
|
-
|
Yes
|
-
|
![ブリスターパックの画像](jpg/img05-4.jpg)
They are in Blister pack except some products.
What is Blister pack?
It is widely used for packing of tables to avoid getting scratched, damp and contamination.